Please use this identifier to cite or link to this item:
http://hdl.handle.net/20.500.12666/839
Title: | Uncertainty in the Calibration Transfer of Solar Irradiance Scale: From Absolute Cavity Radiometers to Standard Pyrheliometers |
Authors: | Balenzategui, José Lorenzo Molero, María Silva, José Pedro Fabero, Fernando Cuenca, José Mejuto, Eduardo De Lucas Veguillas, Javier |
Keywords: | Solar irradiance;Calibration;Cavity radiometers;Pyrheliometers;Solar metrology;Uncertainty;Traceability |
Issue Date: | 2-Apr-2022 |
Publisher: | Multidisciplinary Digital Publishing Institute (MDPI) |
DOI: | https://doi.org/10.3390/solar2020010 |
Published version: | https://www.mdpi.com/2673-9941/2/2/10 |
Citation: | Solar 2(2): 158-185. 2022 |
Abstract: | In this work, the method for calculation of uncertainty of pyrheliometers’ responsivity during their outdoor calibration process in the laboratory is exposed. It is applied first for calibration of standard pyrheliometers by comparison to cavity radiometers, and after for calibration of an end-user pyrheliometer against that standard pyrheliometer. The dissemination of the WRR irradiance scale is illustrated in practice and the increasing uncertainty in the traceability chain is quantified. The way of getting traceability to both WRR scale and to SI units in the current situation, where the shift between these radiometric scales is pending to be solved, is also explained. However, the impact of this gap between scales seems to be more important for calibrations of reference Class A pyrheliometers than in the final determination of DNI irradiance, because in this case, the cumulative uncertainty is large enough as to not significantly be affected for the difference. The way to take into account different correction terms in the measurement model function, and how to compute the corresponding uncertainty, is explained too. The influence of temperature of some pyrheliometers during calibration process and the potential impact on the DNI irradiance calculated with these instruments is exemplified. |
URI: | http://hdl.handle.net/20.500.12666/839 |
E-ISSN: | 2673-9941 |
Appears in Collections: | (Calib.) Artículos |
Files in This Item:
File | Description | Size | Format | |
---|---|---|---|---|
Uncertainty in the calibration transfer of solar irradiance scale, from absolute cavity radiometers to standard pyrheliometers.pdf | 5,02 MB | Adobe PDF | View/Open |
This item is licensed under a Creative Commons License